Product Specific Security Information - Petya and WannaCry

 

Teams at Beckman Coulter have evaluated the risk of the Petya and WannaCry cybersecurity vulnerability in our products. Please see information pertaining to specific products below. If your product is not listed here please contact your local hotline or submit a support request. This page will be updated as additional information becomes available.

Product Line Model Names Recommendations and Comments
Immunoassay Systems Access 2, Unicel DxI 600, Unicel DxI 800 Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance.
Chemistry Systems AU Models (AU480, AU640, AU680, AU 2700, AU5400 & AU5800) Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance. 
Unicel DxC 600, Unicel DxC 800 Microsoft Patch is not applicable. These systems do not use a Windows operating system.
Blood Bank PK7300 Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance. 
Information Systems DxONE Command Central Click here for instructions.
PROService, RAP Box Microsoft Patch is not applicable. These systems do not use a Windows operating system.
DxONE Inventory Manager, RMR Microsoft Patch has been deployed to all customer versions.
REMISOL Click here for configuration based instructions.
Hematology DxH 500 Microsoft Patch is not applicable. These systems do not use a Windows operating system.
DxH 600, DxH 800, DxH SMS Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance. Should your organization require the installation of the Microsoft Patch click here for instructions.
HmX, HmX AL Microsoft Patch is not applicable. These systems do not use a Windows operating system
LH 500 Neither the vulnerability or the Microsoft patch is applicable to this version of Windows.
LH750, LH780, LH785, LH Slidemaker, LH Slidestrainer Neither the vulnerability or the Microsoft patch is applicable to this version of Windows.
CellaVision DM1200, DM96, DM9600 Please contact CellaVision www.cellavision.com
AcT, AcT Diff Microsoft Patch is not applicable. These systems do not use a Windows operating system.
Ac•T 5diff Neither the vulnerability nor the Microsoft patch are applicable to this version of Windows.
Urinalysis iRICELL, IChemVELOCITY, iQ200 Beckman recommended installation is in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance. Should your organization require the installation of the Microsoft Patch click here for instructions.
Lab Automation Power Express Please contact your local service representative for additional information
Aliquoter, Automation Centrifuge, Decapper, Instrument Connections, Outlet, Recapper, Stockyard Microsoft patch is not applicable. These systems do not use a Windows operating system
AutoMate 1200, 1250, 2500, 2550 Click here for instructions.
AutoMate 600/800, LH Auto, LH 1500, LH Custom Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. However, Beckman has successfully validated the patch update. Click here for NH-ISAC guidance.
Power Processor, Power Cell Please contact your local service representative for additional information
PowerLink Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. However, Beckman has successfully validated the patch update.Click here for NH-ISAC guidance.
Ipaw, ISB (Intelligent Sample Banking), Sorting Drive Click here for instructions.
Microbiology LabPro Information Manager
autoSCAN-4
WalkAway-40 SI
WalkAway-96 SI
WalkAway-40 plus
WalkAway-96 plus
Click here for instructions.
Nephelometry IMMAGE, IMMAGE 800 Microsoft patch is not applicable. These systems do not use a Windows operating system.
Molecular VERIS Beckman recommended installation configuration is behind a hardware firewall (RAP box) or in a closed system configuration and is not vulnerable to a network infection. Click here for NH-ISAC guidance. Should your organization require the installation of the Microsoft Patch click here for instructions.
Clinical Centrifugation StatSpin (MP, VT, CytoFuge2, Express2, Express3, Express4, CritSpin Reader) Not Applicable.
The systems do not use or contain the affected software.